Overview of the Test Sets Built-In Tools
Clearing RAM
Clearing RAM
RAM can be cleared using the RAM_MANAGER program in the Test Set’s
ROM. This program clears all RAM, including any SAVE/RECALL
registers saved to Internal RAM.
This program is accessed by running the IB_UTIL program.
Step 1.
Select the SOFTWARE MENU screen.
Step 2.
Select the Select Procedure Location: field and choose ROM.
Step 3.
Select the Select Procedure Filename: field.
Select IB_UTIL from the list of programs. Press Run Test to display
a list of procedures; select RAM_MANAGER and follow directions to clear
RAM.
To preserve the SAVE/RECALL registers, do not use the RAM_MANAGER
program. Instead, load any other test program from a card or from RAM
to clear the previous test program, and then reload the desired test.
See the “Save/Recall” on page 247, and “Saving and Recalling
Instrument Setups” on page 70 for more information about save/recall