Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Connectors
Connectors
Chapter 9
284
AUDIO IN
Two connectors are used to input audio signals to the AF analyzer:
HI is the main audio signal input connection.
LO is used for the audio signal reference. Three choices are available using the AF
ANALYZER screen’s Audio In Lo field:
Gnd connects the center pin through approximately 100 to chassis ground.
Float is a floating input.
600 To Hi provides a 600 internal load to match an audio source with an
output impedance of 600 .
The measured level is the potential between the HI and LO center pins. The shells of
both connectors are at chassis ground.
For more information see “AF Anl In” on page 185, and
“Audio In Lo” on page 190.
Operating Considerations
Input impedance is switchable between 1 M
in parallel with 95 pF, or 600
floating.
This port is selected as the AF analyzer’s input using the AF Anl In field on the AF
ANALYZER screen.
Signals input to the AF analyzer are routed through different filters, amplifiers, and
detectors that affect the displayed measurement.
CAUTION The maximum level between the HI and LO center pins is 42 V peak (approximately 30