Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 7 235
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
(Gen)-(Anl)
This field is used with the RF Offset field to specify the amount of
frequency offset between the RF generator and RF analyzer.
This field is always displayed when the RF Display field is set to Freq.
See “Setting an RF Generator/Analyzer Offset” on page 69 for more
information.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
(Gen)-(Anl) (User Defined)
This field defines the receiver-transmitter frequency offset when using
user-defined channel operation.
Use a positive value (such as 45 MHz) when the radio’s receive
frequency is higher than the transmit frequency.
Use a negative value (such as 45 MHz) when the radio’s receive
frequency is lower than the transmit frequency.
Operating Considerations
The value of this field is only used if the RF Display field is set to Chan,
and the RF Chan Std field is set to USER-DEF.
For more information see the following related topics:
“Chan Space (User Defined)” on page 229
“RF Chan Std” on page 244
“RF Display” on page 245
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE