Chapter 3 61
Operating Overview
Calibrating the Channel Power Measurement – CDMA Analyzer
Calibrating the Channel Power Measurement
– CDMA Analyzer
To calibrate a channel power measurement using the internal CDMA
generator:
Step 4.
Remove power from the RF IN/OUT (or ANT IN) port.
Step 5.
Go to the CDMA ANALYZER screen and choose the Chan Pwr
measurement.
Step 6.
Set the Tune Freq field (or RF Channel field) to the center frequency of
your CDMA signal.
Step 7.
Select the Chn Pwr Cal field and wait until the calibration is complete.