Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 7 245
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
RF Display
This field selects the format for entering the RF generator’s and RF
analyzer’s frequencies:
When Freq is selected, you enter the RF generator’s and the RF
analyzer’s frequencies directly using the keypad or knob.
When Chan is selected, the RF Gen Freq and Tune Freq fields on all
screens are replaced by the RF Channel field, and only the channel
number is entered and displayed.
Channel tuning eliminates the need to enter transmit and receive
frequencies directly into the Test Set. Once your radio’s RF channel
standard is selected, you only have to enter the channel number to
automatically set the RF generator and RF analyzer to the correct
frequencies.
Operating Considerations
See the “RF Chan Std” on page 244 for more information.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
RF Gen Volts
This field specifies whether you want RF voltages expressed as the
voltage across a 50 ohm load, or the open circuit voltage (emf).
Operating Considerations
This setting affects the RF generator’s and the tracking generator’s
amplitudes.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE