Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
174 Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Trig Event
This field determines when the trigger event to the analyzer occurs.
When the selected trigger in this field is qualified by the selection in the
Qual Event field, the analyzer is triggered.
Choices
27 ms is an internally generated clock.
20 ms is an internally generated clock.
80 ms is an internally generated clock.
2 s is an internally generated clock.
Delay controls a trigger delay timer, beginning from the positive
edge of the selected qualifier. A delay of 20 µs to 10 s is valid.
Initially, the time delay will be 100 µs.
Immed triggers immediately after receiving the Qual Event signal.
Default selection: Immed.
Screens on which this field is present
CDMA ANALYZER
COD DOM (Code Domain Analyzer): trigger controls menu
Tune Freq
Use this field to enter the center frequency of the CDMA channel you
are analyzing. This field is displayed only when the RF Display field on
the INSTRUMENT CONFIGURE screen is set to Freq.
NOTE
Channel tuning
You can configure the Test Set to allow direct entry of channel numbers
(instead of entering the frequency) by using the RF Display field on the
INSTRUMENT CONFIGURE screen. See “RF Channel” on page 170
for more information.
Screens on which this field is present
CDMA ANALYZER
COD DOM (Code Domain Analyzer): main controls menu