Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 3 53
Operating Overview
Displaying CDMA Measurements
Displaying CDMA Measurements
Channel Power, Adjacent Channel Power (ACP), Rho,
Error Vector Magnitude (EVM), Average Power
You can select channel power, ACP, rho, EVM, or average power on the
CDMA ANALYZER or CDMA GENERATOR screen as shown in
Figure 3-1. For more information about these measurements see
“CDMA Measurements” on page 131.
Figure 3-1 Displaying Channel Power, ACP, Rho, EVM, or Average Power
ADC Full Scale (ADC FS)
When you select Avg Pwr or Chan Pwr, the ADC FS measurement is also
displayed.
Time Offset, Frequency Error, PN Offset, Carrier Feedthrough
When you select Rho, the Time Offset, Freq Err or PN Offset, (see
Figure 3-1) and Carrier Feedthru measurements are displayed. These
measurements are also displayed on the CODE DOM screen.
Figure 3-2 Frequency Error, and PN Offset
cdmameas pc