Chapter 3 63
Operating Overview
Calibrating the Adjacent Channel Power (ACP) Measurement
Calibrating the Adjacent Channel Power
(ACP) Measurement
To calibrate an ACP measurement using the internal CDMA generator:
Step 12.
Remove power from the RF IN/OUT (or ANT IN) port.
Step 13.
Go to the CDMA ANALYZER screen and choose the ACP measurement.
Step 14.
Set the Tune Freq field (or RF Channel field) to the center frequency of
your CDMA signal.
Step 15.
Select the ACP Cal field and wait until the calibration is complete.