Error Messages
General Information About Error Messages
Text Only Error Messages
Operation errors generally occur when you try to do something the Test
Set was not designed to do. Most messages tell you what to do to correct
the problem, (turn something off, reduce a field’s value, press a certain
key,…and so forth).
Some common messages are listed here:
All self tests passed.
The Test Set did not detect any hardware or firmware failures
during its initial self-diagnostics
This message should always be displayed immediately after
instrument turn on.
Chan Pwr UNCAL: No Freq or Ch Pwr Fltr cal. Start Chan Pwr Cal.
These messages are displayed when a channel power measurement
is being performed but no valid calibration data exists. The message
is accompanied by an audible tone (beep). Remove power from the
ANT IN or RF IN/OUT connector and select Calibrate under the Ch
Pwr Cal field on the CDMA ANALYZER screen. Then reconnect the
input signal to proceed.
ACP UNCAL: No Freq or ACP Offset cal. Start ACP Cal.
These messages are displayed when a channel power measurement
is being performed but no valid calibration data exists. The message
is accompanied by an audible tone (beep). Remove power from the
ANT IN or RF IN/OUT connector and select Calibrate under the
ACP Cal field on the CDMA ANALYZER screen. Then reconnect the
input signal to proceed.
Tune freq out of range for ACP. Change freq or ACP Offset.
This message is displayed if you try to perform an ACP
measurement or calibration that would cause the Test Set to be
tuned to a frequency out of its valid range (for example, the range
between 1015 MHz and 1700 MHz). Note that although the tune
frequency or channel cannot be set to these frequencies, it could be
possible that an invalid tune frequency would be within the range of
an attempted adjacent channel power measurement (offset and BW).
The message is accompanied by an audible tone (beep).
Remove power at ANT IN and RF IN/OUT. Restart calibration.
This message is displayed during channel power or ACP calibration
if a signal is detected at the ANT IN or RF IN/OUT port with an
amplitude high enough degrade the calibration. Remove power at
the ANT IN or RF IN/OUT port, and move the cursor (turn the knob)
to remove the message. The message is accompanied by an audible
tone (beep).