Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 6 193
Analog Measurements - Screens and Control Fields
Control Fields for Analog Measurements
Detector
This setting selects the type of detector used when measuring and
displaying AF signal levels.
Detector Types
RMS displays the root mean square (rms) value of signals.
RMS*SQRT2 displays the rms value of a signal multiplied by the
square root of 2.
Pk+ displays the positive peak value.
Pk displays the negative peak value.
Pk± /2 adds the positive and negative peak values, and divides the
sum by 2.
Pk± Max compares the positive and negative peaks and displays the
greater value (polarity is not indicated).
Pk+ Hold displays and holds the positive peak value until the
measurement is reset. To reset, use
Meas Reset
(press and release
the
Shift
key, then the
Hold
key), select a different detector, or
re-select the same detector.
Pk Hold displays and holds the negative peak value until the
measurement is reset. To reset, use
Meas Reset
(press and release
the
Shift
key, then the
Hold
key), select a different detector, or
re-select the same detector.
Pk±/2 Hold divides the sum of the positive and negative peak values
by 2, and displays the value until the measurement is reset. To reset,
use
Meas Reset
(press and release the
Shift
key, then the
Hold
key),
select a different detector, or re-select the same detector.
Pk± Mx Hold compares the positive and negative peaks and displays
the greater value until the measurement is reset. To reset, use
Meas
Reset
(press and release the
Shift
key, then the
Hold
key), select a
different detector, or re-select the same detector.
Screens on which this field is displayed
AF ANALYZER
Display
This field selects which limits to display on the spectrum analyzer’s
mask: Upper, Lower, Both, or Off.
Screens on which this field is displayed
SPEC ANL: mask controls menu