Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 7 247
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
RF Offset
This field is displayed when the RF Display field is set to Freq.
This field enables/disables the RF generator RF analyzer frequency
offset specified in the (Gen)-(Anl) field below it.
Operating Considerations
When an RF offset is used, changing the RF generator’s frequency or
RF analyzer’s tune frequency automatically alters the other setting.
See “Setting an RF Generator/Analyzer Offset” on page 69 for more
information.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Save/Recall
This field specifies which memory device the Test Set accesses when the
save and recall functions are used.
Internal is a section of internal RAM. RAM is also used for running
IBASIC programs, which may require you to delete the save/recall
registers if the program is very large.
Card is the front-panel MEMORY CARD (PC card) slot. A
Save/Recall Device is not Present message is displayed if you
try to save or recall an instrument setup when a write-able PC card
is not installed in the Test Set.
RAM refers to RAM disks that you can create on internal RAM. Refer
to the Programmer’s Guide for information on creating RAM disks.
This is part of the same memory used when “internal” is specified,
and may have to be erased when loading very large IBASIC
programs. A Save/Recall Device is not initialized message is
displayed if you try to save or recall an instrument setup when a
RAM disk has not been created.
Screen(s) Where Field is Present
I/O CONFIGURE