Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
Chapter 7 227
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
Control Fields on the Configuration Screens
Abort Print
Select this field to interrupt the printing job in progress.
Abort Print is not directly programmable over the GPIB.
Screen(s) Where Field is Present
PRINTER CONFIGURE
Antenna In
This field is used to indicate losses or gains between the ANT IN port
and the device-under-test.
Enter a positive value to indicate a gain (such as an amplifier). The
spectrum analyzer’s marker level (Lvl) measurement is automatically
reduced by that amount. The spectrum analyzer’s Ref Level is
automatically decreased by the same amount, so the trace position does
not appear to change.
Enter a negative value to indicate a loss (such as cable loss). The
spectrum analyzer marker’s level (Lvl) measurement is automatically
increased by that amount. The spectrum analyzer’s Ref Level is
automatically increased by the same amount, so the trace position does
not appear to change.
This field is only used when the RF Level Offset field is set to On. See
“RF Level Offset” on page 246.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE