Agilent Technologies 8935 series e6380a Cell Phone Accessories User Manual


 
240 Chapter 7
C:\Spk\Ref\RefGuideRevE\Cnfgscrn.fm
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
Notch Coupl
This field selects if the Notch Freq setting of the AF ANALYZER
screen is coupled to the AF Gen1 Freq setting. When set to None, the
notch filter and AF generator 1 do not interact. When set to AFGen1
(coupled), the settings track each other unless the AF generator’s
frequency is set outside the 300 Hz to 10 kHz limits of the notch filter.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Number to Call
This field is used enter the dialing sequence to be used whenever the
modem originates a call. The maximum number of characters is 36.
This setting is retained when the instrument is turned off.
Screen(s) Where Field is Present
MODEM CONFIGURE
Opt CDMA TB
This control is provided for future enhancements. It must be set to
Internal until options are available.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Parity
This field specifies the serial communication parity setting when using
the serial ports.
This setting is retained when the instrument is turned off.
Operating Considerations
When in remote mode (Remote Mode field is set to On), this field is
inaccessible.
Screen(s) Where Field is Present
I/O CONFIGURE