4.5.3.4 F3: Setting the reference value
Set the reference value and/or scale value here. If Copying the target value to the reference
value has been specified in the basic setup, the setup guidance for the reference value will
not be displayed, however, setting the scale value is permitted.
In this example assume that the reference value is 12.5 mm, and the scale value is 1.
PROG
Step 1: The previously set reference value is displayed.
Enter 12.5.
1
PROG
If 1 is entered the currently displayed
setup value changes to 1, however, it will
not be saved in memory until the
ENT
key is pressed.
2
PROG
Enter 2.
.
PROG
Enter a decimal point (.).
5
PROG
Enter 5.
PROG
Step 2: Press the
ENT
key.
The setup data will be saved in memory and the
operation automatically proceeds to the scale
value setting.
For information about the relationship between the
analog voltage output and scale value, refer to
Section 4.5.3.5, Analog voltage output and scale
value.
PROG
Enter a scale value of 1.
PROG
Step 3: Press the
ENT
key.
The setup data will be saved in memory and the
operation automatically proceeds to the offset
value setting.