a) Single measurement
One session of measurement will be performed according to the one specified program.
The factory default setting is this single measurement.
b) Simultaneous measurement
• In one measurement session two specific programs are executed at one time as a pair.
Combinations of program numbers to form these pairs are shown in the following table.
• To run a pair of programs, either of the two can be specified via one of the numeric
keys and the one specified is called “foreground program”, and its counterpart is called
“background program”.
<<Possible combinations in the 100 program mode>>
Program
numbers pairs
Channel No. (Digit of ten of the program number) [0 to 4]
01234
00 & 05 10 & 15 20 & 25 30 & 35 40 & 45
01 & 06 11 & 16 21 & 26 31 & 36 41 & 46
02 & 07 12 & 17 22 & 27 32 & 37 42 & 47
03 & 08 13 & 18 23 & 28 33 & 38 43 & 48
04 & 09 14 & 19 24 & 29 34 & 39 44 & 49
Program
numbers pairs
Channel No. (Digit of ten of the program number) [5 to 9]
56789
50 & 55 60 & 65 70 & 75 80 & 85 90 & 95
51 & 56 61 & 66 71 & 76 81 & 86 91 & 96
52 & 57 62 & 67 72 & 77 82 & 87 92 & 97
53 & 58 63 & 68 73 & 78 83 & 88 93 & 98
54 & 59 64 & 69 74 & 79 84 & 89 94 & 99
<< Possible combinations in the 10 program mode>>
Program
numbers pairs
00 & 05
01 & 06
02 & 07
03 & 08
04 & 09
3.2.2 Basic setup
• This is used to customize the basic setup conditions, including the resolution, available
functions, and interface conditions, according to the measurement requirements.
• This basic setup must be performed at the beginning of a measurement. Note that changing
the setup of resolution, simultaneous measurement, or dual-unit measurement in this basic
setup cancel the existing calibration values and function setup.
• The basic setup mode is entered by turning on the power while holding down the
SET
key.
Note that no response will be made to an I/O interface input and RS-232C/GP-IB com-
mand in the basic setup mode.
• For more information, refer to Section 4.1, “Basic Setup”.