A SERVICE OF

logo

6 - 18
No. 99MBC095A
(5) SG sssssss, SG aaa-bbb
(a) Format: SG sssssss (sssssss: SEG No. Number of digits should be between 1 and 7.
Duplicated number must not be specified.)
SG aaa-bbb (aaa: start edge, bbb: finish edge. The range is between 1 and 255
for both edges. However, aaa should not be identical to bbb.)
(b) Description: Setting the segment (measuring position).
Two types of setting are available; segment specification and edge specification.
(c) Example: Reception command SG2 Response command OK
Reception command SG2-65 Response command OK
(d) Supplement: Segments and edges should be set in the basic setup.
sssssss can be set with 7 digits or less.
Ex.) SG 1234567, SG 24, SG3, etc.
aaa and bbb should be set within 3 digits.
Ex.) SG1-2, SG 2-33, SG 111-255, etc.
The order of the start edge and finish edge can be reversed.
(6) MN
(a) Format: MN nnnn (nnnn: Number of scans, between 1 and 2048)
(b) Description: Set the averaging method to the arithmetical average, and specify the number
of scans to nnnn (2
n
, where n= 0 to 11).
(c) Example: Reception command MN 1024
Response command OK
NOTE The number of scans must be between 16 and 2048 if the ultra-fine wire measure-
ment is specified in the basic setup.
(7) MNH
(a) Format: MNH nnnn(nnnn: Number of scans, between 32 and 2048)
(b) Description: Set the averaging method to the moving average, and specify the number of
scans with nnnn. nnnn is 2
n
, where n= 5 to 11.
(c) Example: Reception command MNH 1024
Response command OK
(d) Supplement: MNL command is as same as MNH command
(8) STR
(a) Format: STR
(b) Description: Data that has been set by the RS-232C command will be erased from memory
if the power is off. To retain the data after the power off, use this command to
save the critical measuring conditions in memory. But the ST and NST
command will not be saved.
(c) Example: Reception command STR
Response command OK
(9) LOCK
(a) Format: LOCK
(b) Description: Locks the keyboard of this machine to prevent accidental operation.
To release this key lock state, use the UNLOCK command.
(c) Example: Reception command LOCK
Response command OK
(d) Supplement: Lock set by this command can not be released with key operation.