4 - 21
No. 99MBC095A
4. SETTING UP THE MEASURING CONDITIONS
4.1.2.8 Setting in the B7 mode (expanded items)
To use this mode, turn on the power while pressing the
9
key. Items in this mode will be
displayed for selection after the display of the basic setup items in the B6 mode.
IMPORTANT • If expanded items (of the mode of B7) that are not required are displayed, cancel
the display by setting in place of for the expanded item setup.
• For setting in the space for additional function always select which is the
default or . If this setting is carelessly modified, this Unit will operate
unexpectedly such as to disable any measurement.
a. Setting expanded items (Guidance: )
Set whether to use (display) the function of expanded items.
: No use (display) of the expanded item
: Use (display) of the expanded item
(Default setting: )
PROG
Step 1: Setting changes between and at
each entry of .
Step 2: Press the
ENT
key while the setting is flashing.
When this entry is made, the next setting item will
be displayed.
b. Reservation (Guidance: )
PROG
(Default setting: )
c. Reservation (Guidance: )
PROG
(Default setting: )
d. Reservation (Guidance: )
PROG
(Default setting: )
e. SHL setting (Guidance: )
Sets up the SHL value for the measurement of, for
example, the transparent sheet width. To use this
function, set for "a. Expanded item" and
for "g. Setting measurement mode". For detailed
information about this setting, refer to Section 3.2.4.1,
"Transparent object".
Be aware that any change in the parameter setting greatly affects the measured data
(measuring accuracy).
(Default value: (%))
Setting value is an integer between 5% and 95%.