6 - 6
No. 99MBC095A
• Output in the ready state (if the basic setup is determined to enable output in the ready
state)
T
SO
RUN
RES
ACK
STB
GO, ±NG
T
DO
T
SO
T
DO
HOLD
T
IN
or more
• Response time
Item Response time Description
Input time
Acceptance time
Delay time by filter
Duration of Single-run measurement
Duration of continuous-run measurement
Duration of ready state
T
IN
T
AD
T
if
T
SO
T
BO
T
BC
T
BD
> (Tif x 2 + 3 ms)
< (Tif x 2 + 3 ms)
5 ms, 2 ms, 20 ms
Can be selected by Expanded basic setup (“7 IFF”).
Refer to p.6-37, p.6-38, Strobe length (T
SO
).
Data setup time
Strobe length
T
DO
0.05 ms to 0.2 ms
< (Tmr x N + 3 ms)
< (Tmr x N + 1 ms)
< (Tmr + 1 ms)
Measurement intervalT
mr
Refer to p.6-37, p.6-38, Measurement interval (Tmr)
Number of samples (SMP N)
N
Refer to Section 9.2.4, Number of samples (“6 SMP N”).
And refer to Section 4.5.3.8, Number of samples (“6 SMP N”)
• Other
1. Use negative-true logic pulses of TIN or more for the input signals.
2. RES signal clears the previous measurement result and interrupts the measuring
operation.
3. Simultaneous input of multiple signals is not accepted.
4. During measurement only RES, RUN or HOLD signals are accepted.
NOTE 1. Zero-run measurement is enabled only if the RUN input is set to “S.RUN” in the
basic setup.
2. While the HOLD signal is ON, the GO, ±NG, STB and analog output signals are
held without being updated.