5 - 14
No. 99MBC095A
3. Applied measurement of preset 3
Measuring unit 1
1
5
D
Measuring unit 2
[figure c]
Make the combined preset enabled in the DF-type dual-
unit measurement.
Example of [figure c]
Suppose that D = 250.0
±0.05
mm.
Basic setup
Set the dual-unit measurement type to DF type.
Function setup
1. Segment = (1 + 5)
2. Number of scans for averaging = 512 times
3. GO/NG judgment
a. Lower limit value = 249.95 mm
b. Upper limit value = 250.05 mm
4. Preset value
a. Set to 250.0 mm.
b. Direction = 1 (negative)
Measurement
PROG
BUSY
RUN
+NG
GO
-NG
Before setting the preset function, the display appears
as shown at the right.
Since the preset function is not enabled yet, the gap
(segments 1 +5) will be measured.
PROG
BUSY
RUN
+NG
GO
-NG
Set the preset value to 250.0 mm. Since this is a gap
measurement, set the direction to 1 (negative).
PROG
BUSY
RUN
+NG
GO
-NG
Press the
DATA C/RUN
key.
will be displayed and single-run
measurement will start.
After the set measurement interval the measured value
is latched on the display, and the GO/NG judgment
result is outputted.
NOTE About the combined preset
1. The combined preset is used to measure a workpiece that has a dimension close
to that of the reference gage. For measuring workpieces with various dimensions,
use the combined calibration.
2. To use the combined preset together with the combined calibration, set the
direction to 0 (positive).
3. If the combined calibration or individual calibration for each Measuring Unit is
canceled, the combined preset is also canceled.