6 - 22
No. 99MBC095A
(19) MC, MCAL
(a) Format: MC
MCAL
(b) Description: Both the MC and MCAL commands are used to clear the statistical memory.
This operation is required before starting statistical processing.
(c) Example: Reception command MC or MCAL
Response command OK
(20) STAT
(a) Format: STAT
(b) Description: Requests the statistical processing data. The statistical processing data will be
cleared when the power is off.
(c) Example: Reception command STAT
Response command STAT
DATA, P00, N100, AVG12.0001,
MAX12.0005, MIN11.9998, RNG0.0007, S.D0.00007
(21) AUT
(a) Format: AUT, Nmmm, Dtttt, L±ddd.dddd, H±ddd.dddd
(b) Description: Set the conditions of automatic workpiece detection with the following data to
follow AUT and delimited by a comma (,).
Lower and upper detection limits for the position detection method do not
require a - sign, so it will be ignored if specified.
Responds only when the automatic workpiece detection has been set in the
basic setup.
N mmm (mmm: number of measurement times between 1 and 999. If 0 is
specified, automatic workpiece detection is not performed.)
Dtttt (tttt: Invalidation period between 0 and 9999 ms)
L±ddd.dddd (±ddd.dddd: Lower detection limit)
H±ddd.dddd (±ddd.dddd: Upper detection limit)
(c) Example: Reception command AUT, N50, D15, L9.5, H12.3
Response command OK
(22) AUT, S
(a) Format: AUT, S
(b) Description: Where Performing the automatic workpiece detection is specified in the basic
setup, and if this command is received, S will be responded each time a
workpiece is detected.
If this setup is not made in the basic setup, designating this command results
in ER6.
(c) Example: Reception command AUT, S
Response command OK
(23) RA
(a) Format: RA
(b) Description: Transmits as the response the conditions of the automatic workpiece detection
using RA FORMAT.
(c) Example: Reception command RA
Response command AUT, N50, D15, L9.5, H12.3