![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/f/95/f951be8e-3276-4696-92fa-ccb64a2c46c9/f951be8e-3276-4696-92fa-ccb64a2c46c9-bg50.png)
4 - 15
No. 99MBC095A
4. SETTING UP THE MEASURING CONDITIONS
c. Setting the number of scans (Guidance: )
Set the number of scans that are used for the automatic workpiece detecting function.
: Detection from 16 scans
: Detection from a single scan
(Default setting:
)
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and . While
the desired setup option is flashing, press the
ENT
key. The operation automatically enters the
process for setting the group judgment.
NOTE Even if 16 scans are specified in the position detection method, the actual detecting
operation will be performed with a single scan.
d. Setting the group judgment (Guidance: )
Set whether to use the group judgment function.
: Does not use the group judgment function.
: Uses the group judgment function.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and .
While the desired setup option is flashing, press
the
ENT
key. If is selected, the opera-
tion proceeds to B4: Setting the use of RS-232C
baud rate, and if is selected, the operation
enters the process for setting the group judgment
result output function.
e. Setting the group judgement output (Guidance: )
Set whether to output the group judgment result to the RS-232C/GP-IB interface.
: Does not output the group judgment result to the RS-232C/GP-IB interface.
: Outputs the group judgment result to the RS-232C/GP-IB interface.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and .
While the desired setup option is flashing, press
the
ENT
key. The operation automatically
proceeds to B4: Setting the use of RS-232C port.