A SERVICE OF

logo

4 - 14
No. 99MBC095A
4.1.2.4 Selecting and setting the function in the B3 mode
a. Setting the abnormal value elimination function (Guidance: )
Set whether to use the abnormal value elimination function.
: Does not use the abnormal value elimination function.
: Use the abnormal value elimination function.
If the number of samples has been set, measurement will be finished when the
measured data within the limit value are obtained for the specified sample
number, and calculation is performed only with the within-the-limit data for
displaying the measurement result.
: Use the abnormal value elimination function.
If the number of samples has been set, measurement will be performed for the
specified sample number and displays the measurement result after a calcula-
tion has been performed only with the within-the-limit data. The message
Err-0 will be displayed if no data is obtained within the limit value.
(Default setting:
)
PROG
Step 1: Each time the key is pressed the displayed
setup option changes in following order :
.
While the desired setup option is flashing, press
the
ENT
key. The operation automatically enters
the process for setting the automatic workpiece
detecting function.
b. Setting the automatic workpiece detecting function (Guidance: )
Set whether to use the automatic workpiece detecting function.
: Does not use the automatic workpiece detecting function.
: Performs automatic workpiece detection with the diameter detection
method.
: Performs automatic workpiece detection with the position detection
method.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option changes in the following order:
. While the
desired setup option is flashing, press the
ENT
key. If (the automatic workpiece
detecting function is not used) is selected, the
operation proceeds to setting the group judgment,
otherwise it enters the process for setting the
number of scans.