3 - 11
No. 99MBC095A
3. DISPLAYS AND KEY OPERATIONS
3.2.3 Function setup
Use this procedure to set up the conditions necessary for measurement.]
For each program number register measurement conditions including the segment (part
feature to be measured), measurement interval (measurement time), and GO/NG judgment
criteria that are the best suited for the objective workpiece.
To enter the function setup mode press the
SET
key in the ready state. Each of the
LIMIT
,
SHIFT
P.S V/
P.SET
,
MASTER/REF
,
SHIFT
MASTER/REF
keys allows the individual
setup item to be established, and the
key enters the setup operation for items which
are most frequently accessed for set up.
For more information refer to Section 4.5, Setting Up the Functions.
3.2.4 Setups according to the property of each workpiece
For measuring workpieces that transmit light or have a dimension smaller than the diameter
of the scanning beam it is critical to make setups that take into account the properties of the
workpiece.
3.2.4.1 Transparent object (Workpiece that transmits light)
a) Round bar
Workpieces such as fiber optics and glass tubes are more or less transparent, while
workpieces made of steel are not. This requires different segment settings.
The segment settings for an opaque object and a transparent object are as follows:
Setup for measurement of transparent or opaque object is possible in the basic setup.
Segment 2
Segment 4
Segment 1Segment 1
Segment 3 Segment 2
Segment 5 Segment 3
Transparent
Workpiece
For transparent object
Laser scan direction
Photo-electric signal
For opaque object
Binary voltage (SHL)