No. 99MBC095A
3) Installing the interface
If installing more than two optional interfaces, install them in the order below:
a. Digimatic Output Unit
b. Second Analog I/O, BCD, or GP-IB interface
c. Dual-type add-on unit
Remove the protection plate from where each interface is to be installed.
Install each interface unit inside the display unit following 6.3.1 through 6.3.2.
Be sure to tighten the fixing screws of each interface unit.
4) Replacing the cover
Replace the cover which has been removed at step (2) and fit it with four screws.
6.3.1 Digimatic Output Unit
Using the provided connecting cable, connect the Digimatic Output Unit to connector
CN14.
Fit the connecting cable to the cut on the connector CN14 and firmly connect them.
Firmly tighten the Digimatic Output Unit mount plate with two screws as shown.
10 1
CN14
Digimatic Output Unit
CN14
Connect the connecting cable to CN14.
Cut on CN14
6.3.2 Second Analog I/O, BCD, and GP-IB interfaces
Using the provided connecting cable, connect the interface unit to connector CN9.
Fit the connecting cable to the cut on the connector CN9 and firmly connect them.
Firmly tighten the Interface Unit mount plate with two screws as shown.
CN9
34
33
2
1
Connect the connecting cable to CN9
Cut on CN9CN9
Interface unit