2 - 4
No. 99MBC095A
Terminal to which a grounding wire is connected.
Ground the system using the provided grounding wire to avoid the effect of interfer-
ence noise caused in the setup environment.
Step 6: Checking the remote interlock connector
Make sure that the short-circuiting pin is inserted into the REMOTE INTERLOCK
connector on the rear panel of the Display Unit. If this short-circuiting pin is not
inserted, laser emission is disabled, even if the power switch is on.
To emergency stop laser emission, refer to the following diagram.
Switch ON: Laser emission ON
Switch OFF: Laser emission OFF
(
Manufacturer: Sato Parts
)
Applicable connector: PJ-2
5V, 3mA
Short-circuiting pin
Switch
NOTE Recovering operation is not guaranteed. Never use this function for other than
emergency stop.
Step 7: Connecting the interface
For information about the procedure used to connect the interface, refer to Section
6.1.1, I/O Analog Interface and Section 6.1.2, RS-232C Interface.
NOTE 1. Note the following when connecting the signal cable.
For information about the precautions to be observed when connecting the signal
cable refer to the measuring unit users manual.
2. Note the following when making cable connections.
Always make connection or disconnection with the power cord unplugged. In
addition, before connecting to the interface make sure that the power to all other
units connected or to be connected are also off.