No. 99MBC095A
4. SETTING UP THE MEASURING CONDITIONS
e. Setting the DXY-type calculation (Guidance: )
Set the method for calculating the measurements obtained from two Measuring Units.
: Calculates a sum, (X+Y).
: Calculates a mean, (X+Y) /2.
: Calculates a difference, (X-Y).
: Calculates half of the difference, (X-Y) /2.
(Default setting:
)
PROG
Step 1: Each time the key is pressed the displayed
setup option changes in the following order:
→ → →
. While the desired setup option is
flashing, press the
ENT
key. The operation
automatically proceeds to B3: Setting the abnor-
mal value elimination function.
f. Selecting the method of specifying segments (Guidance: )
Select the method of specifying the measurement position from segment specification and
edge specification.
: Uses segment specification.
: Uses edge specification.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and
. While the desired setup option is
flashing, press the
ENT
key. The operation
automatically enters B3: Setting the abnormal
value elimination function.
NOTE If any of the following setting is performed, the system automatically proceeds to the
stage of segment setup without displaying the guidance for the
method of specifying segments SEG:
1. a. is selected in Setting the workpiece type.
2. b. is selected in setting whether to perform ultra-fine wire measurement.
3. d. or is selected in setting the dual-unit measurement