6 - 20
No. 99MBC095A
Number of sample measurements , SMP nnn
Sample measurement • Maximum value *6 , MAX
• Minimum value *6 , MIN
• Range *6 , RNG
• Mean *6 , AVG
Group size subject to judgment *2 , GN nn
Lower tolerance limit of group judgment *2 , GLL±ddd.dddd
Upper tolerance limit of group judgment *2 , GLH±ddd.dddd
Statistical items for group judgment: Maximum value *2, 6 , GMX
: Minimum value *2, 6 , GMN
: Mean *2, 6 , GAG
: Range *2, 6 , GRG
(b) Description: This sets the measuring conditions.
• Each of the commands that follow the SET command must be delimited by a
comma (,).
• A command which doesn’t need a setting change can be eliminated.
• Approximately 0.5 second is required for this command to be processed.
(c) Example: Reception command SET, SG2, M4, LL
∆ 5.988, LH∆ 6.010, REF0
*1: Select either setup method in the basic setup.
*2: This is valid only if the function is specified in the basic setup.
*3: This is valid only if the “Copying the target value to the reference value” is specified in
the basic setup.
*4: Set the reference gage on the Measuring Unit before sending this command. These
commands will spend several seconds for processing. The settings of each function are
as follows:
• Positive preset: Presetting in the positive (0) direction.
• Negative preset: Presetting in the negative (1) direction.
*5: “PRr” is used to set the data output conditions for the RS-232C (printer)/GP-IB or
Digimatic Output Unit. If the PR number is 1, 3 or 5, it is possible to set the periodic
output timer, and the data output interval can be selected from 0 (for each measure-
ment) and between 1 and 999 seconds.
The PR numbers and the data output conditions have the following relationships.
Data output condition
(PR No.)
Printer
RS-232C
GP-IB
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
—
—
—
—
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputs data for each measurement.
: Performs measurement and outputs data when a GO measurement results.
: Performs measurement and outputs data when a ±NG measurement results.
: No output
—
—
—
*6: Only one of these statistical items can be specified.