6 - 35
No. 99MBC095A
6. INTERFACE UNIT
7) Example for PLC (Programmable Logic sequence Controller)
24V DC Power supply
COM (IN)
2
ACK 33
STB 34
RUN 7
RESET 25
GO 15
+COM (IN)
–COM (OUT)
INPUT
INPUT
+
INPUT
OUTPUT
OUTPUT
PLC
COM (OUT)
35
LSM 2nd I/O
The input of PLC must be plus common type.
The output of PLC must be minus common type.
There must be Transistor type input or output. Do not use Reray type.
8) Example for High sensitivity relay
24V DC
Power supply
NG
32
K1
Lamp
or
Buzzer
COM (OUT)
D1: 10EDA10 (Nihon Inter Electronics)
35
K1
+NG
+
+
14
D1
TIP When Err-0 occurs, judgment output are all set to OFF.
The single-run measurement and zero-run measurement will be interrupted by Err-
0 and the measurement mode will be terminated without outputting the results.
During the continuous-run measurement, the ACK is set to ON, even if an Err-0
occur. Measurement continues after the error is rectified.