A SERVICE OF

logo

3 - 19
No. 99MBC095A
3. DISPLAYS AND KEY OPERATIONS
3.2.6.2 DXY type
Used to perform X-Y (2-axis) measurement with two Measuring Units positioned perpen-
dicular to each other.
For DXY-type measurement the objective portion of the workpiece to be measured should
be selected through segment specification, not by edge specification.
Select segments 1 to 3 for Measuring Unit 1, and select segments 5 to 7 for Measuring
Unit 2.
Measuring unit 1
Measuring unit 2
(Y-axis direction)
Display unit
(X-axis direction)
SEG 1
SEG 5
SEG 2
SEG 3
SEG 6
SEG 7
Least significant digit of
program number: 0 - 4
Least significant digit of
program number: 5 - 9
Calibration
Select each segment (between 1 and 3) for Measuring Unit 1 with the least significant
digit of program number between 0 and 4, and select each segment (between 5 and 7) for
Measuring Unit 2 with the least significant digit of program number between 5 and 9.
Any program number can be assigned to segments 1 to 3 (in the X direction) and segments
5 to 7 (in the Y direction).
If one segment for each of the X and Y directions are selected (e.g. 2 + 6), calculation is
made possible between the measurements from these segments in the X and Y directions.
The calculation items can be selected in the basic setup by arithmetic addition: (X + Y),
mean outside diameter: (X + Y) / 2, Difference: (X-Y), and film thickness: (X-Y) /2.