4. SETTING UP THE MEASURING CONDITIONS
4.2 Calibration
The LSM system can be calibrated quite easily and with high accuracy.
4.2.1 Calibration gages and gage stand
Supported calibration gages and gage stand have the following shapes.
Calibration gages and stand
With-holder type Straight type Stepped type
Gage stand
4.2.2 Entering the calibration mode
Enter the calibration mode with the following procedure.
< Preparation >
(1) Turn on the power and wait at least 30 minutes for the system to thermally stabilize.
(2) Prior to use, wipe dust and oil from the gage and gage stand with a cloth soaked in
alcohol or thinner. If calibration has been completed, carefully store them in a dedicated
case after applying a rust preventive oil to their surfaces.
(3) Specify SEG 2.
For information about the method of segment specification, refer to Section 4.5, “Setting
Up the Functions”.
PROG
On edge specification, select either manual measurement or automatic measurement with
respect to diameter.
PROG PROG
a) Manual measurement b) Automatic measurement
Set the start edge to 2 and the end edge to 3.
PROG PROG
a) Start edge b) End edge