6 - 3
No. 99MBC095A
6. INTERFACE UNIT
(2) Output circuit
51
Output signal
(GO, -NG, +NG, STS)
GND
Maximum rating of the output transistor is 30 V, 50 mA.
(3) Analog output
Refer to Section 6.2.2.4 , “Analog output”.
TIP 1. If Err-0 (specified workpiece not present) occurs, the following remedies are
taken:
±NG and GO signals are turned OFF if this error occurs.
Single-run measurement and zero-run measurement will be terminated without
outputting the results.
During continuous-run measurement or continuous-run measurement with term
specification ACK remains ON. The measurement can be resumed after Err-0 is
rectified.
2. Connecting the second I/O interface will disable the input circuit/output circuit of
the standard interface.
(Analog output can be used.)