![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/f/95/f951be8e-3276-4696-92fa-ccb64a2c46c9/f951be8e-3276-4696-92fa-ccb64a2c46c9-bg5c.png)
4 - 27
No. 99MBC095A
4. SETTING UP THE MEASURING CONDITIONS
PROG
Step 4: If the
ENT
key is pressed to save the HIGH CAL
setup value in memory, the operation automati-
cally returns to the ready state.
Step 5: Set the LOW CAL gage.
As with the HIGH CAL gage, the LOW CAL
gages vary in shape depending on the LSM model
to be calibrated. Set the LOW CAL gage so that
the center of the calibration range is properly
measured.
A LOW CAL gage for calibrating dimensions less
than 2 mm should be set so that it fits with the
mounting hole of the gage stand.
PROG
Step 6: Set up the LOW CAL gage.
In the ready state press the
L.CAL
key. The
previously set LOW CAL value is displayed, and
the LOW CAL setup mode is entered.
Step 7: Enter the verified dimension of the LOW CAL
gage.
PROG
5
00
0
1
.
PROG
Step 8: If the
ENT
key is pressed to save the LOW CAL
setup value in memory, the operation automati-
cally returns to the ready state.
4.2.3 Combined calibration
Measuring unit 1
1
5
Measuring unit 2
Segment
For DF-type setup, it is possible to perform combined
calibration with two Measuring Units.
In the example as shown at the right hand, a workpiece is
mounted so it extends over two Measuring Units and a
reference pins are used to improve the measurement
repeatability.
In this case, a combination of segments (1 + 5) is used.
The method for setting up the calibration value is the same
as that used for setting the HIGH CAL or LOW CAL
gage.