3 - 31
No. 99MBC095A
3. DISPLAYS AND KEY OPERATIONS
3.2.16 Group judgment
• While the tolerance judgment is applied to each measurement from a workpiece, this group
judgment is applied to a group of the specified number of workpieces.
Output of
judgment
result 1
Output of
judgment
result 2
Output of
judgment
result 3
Output of judgment result 1
Output of
judgment result 2
Output of
judgment
result 4
Output of
judgment
result 5
Output of
judgment
result 6
Output of
judgment
result 7
Judgment to Group 1 Judgment to Group 2 Judgment to Group 3
Output of group judgment result
Individual
judgment
1
Individual
judgment
2
Individual
judgment
3
Individual
judgment
4
Individual
judgment
5
Individual
judgment
6
Individual
judgment
7
Individual
judgment
8
Output of individual
judgment result
• In the basic setup select whether group judgment is to be performed. If it is, then set the
group size (the number of workpieces included in a group), calculation items (mean,
maximum value, minimum value, and range), and group lower limit and upper limit. If
“Not performing group judgment” is selected, the setup guidance for it will not be dis-
played.
• The group judgment will be in effect in a single-run measurement or continuous-run
measurement.
a) For the result display and GO/NG judgment indication each individual measurement
and judgment result will be used.
b) Output of judgment result
1. If only the standard Analog I/O Interface is used
Each individual judgment result will be outputted.
2. If the second Analog I/O Interface is used
Each individual judgment result will be outputted for A-(+NG), A-(GO), and A-(-
NG), and the group judgment result will be outputted for B-(+NG), B-(GO), and B-(-
NG), respectively.
c) RS-232C/GP-IB output
In the basic setup it is possible to set whether the group judgment result data is output-
ted for the RS-232C/GP-IB interface. If it is, the output contents from the group
judgment will be as follows:
P0, ( GO) 12.34567 ... Individual data
P0, ( GO) 12.34560 ... Individual data
P0, (+NG) 12.34600 ... Individual data
GP0, ( GO) 12.34575 ... Group judgment result data
• Each individual piece of measurement data can be the objective of statistical processing,
however, group measurement data will be excluded from statistical processing.
• Even if “Err-0” (specified workpiece not present) occurs, the obtained data will not be
cleared. To abort the measurement, press the
C
key (or input RESET via the I/O
Interface or the “CL” command via the RS-232C/GP-IB interface).