6.1.2.6 List of response commands if an error occurs
Response command GP-IB SRQ status byte Description
A workpiece is not present in the specified segment.
• A workpiece is not set properly.
• Shutter is closed.
Any error of communication
• The communication setting of the opponent instrument is different from LSM.
• Check the setup contents in the basic setup.
• Isolate the cables from noise sources.
Remark: “ER6” or “ER9” may occur in the first communication right after the start up. If this happens, repeat the command transmission of
“CL” until the reception of “OK” is confirmed.
A numeric value greatly different from the reference gage dimension is set.
• Limit values for go/no-go judgment and abnormal data exclusion have been set in reverse
order or equal.
• Input value is too large.
An unavailable command is received.
• Command format is incorrect.
• Baud rate and/or data bits are not consistent.
Message from the external device
• Measurement is interrupted by signal input from key operation or I/O interface.
ER0
ER2
ER9
ER5
ER6
ER7
0RSV001000
0RSV001010
0RSV001101
0RSV001110
0RSV001111