A SERVICE OF

logo

2 - 2
No. 99MBC095A
Step 2: Attaching the ID unit
1.Loosen the two screws that secure the ID unit protection cover at the left on the
real panel of the Display Unit and remove the cover by sliding it rightward.
2.Remove the dummy ID unit (amber) that has been mounted at the left of the
“TRANSMITTER-1” connector on the rear panel of the Display Unit, then insert
the ID unit (beige) that comes in the same package as the Measuring Unit.
This ID unit stores critical data that ensures the accuracy of the Measuring Unit
and has the same serial number as the accompanying Measuring Unit. Confirm that
these two numbers are identical before inserting the ID unit.
SCAN SIG.-1ID UNIT 1
TRANSMITTER-1
REMOTE INTERLOCK
ID
ID unit
ID unit slot
Serial number label
ID unit Measuring unit
If a Dual-type Add-on Unit is used, install the ID unit in the ID UNIT 2 slot above
the ID UNIT 1.
3.Replace and secure the ID unit protection cover revering the procedures in step 1
above.