3 - 22
No. 99MBC095A
e1
θ
S
Photoelectric device
Reception lens
Workpiece
e2
θ
S
Reception lens
(b) Measuring a large diameter workpiece
(a) Measuring a small diameter workpiece
Workpiece
If a large-sized workpiece is measured using the DF-type setup, combined calibration
will ensure more accurate measurment.
c) Combined preset
While combined calibration is used to measure workpieces of various sizes, the com-
bined preset is used to measure workpieces with a dimension close to the reference
gauge.
Since, at a combined preset, separate values can be set for each program number, a
maximum of 100 kinds of workpieces can be registered. To perform the combined
preset make the measuring positions of the reference gauge and workpiece as consistent
as possible.
Designate the combined preset by combining segments such as (1 + 5) or (1 + 7), and
set the direction to negative (1).
However, if this is coupled with a combined calibration, set the direction to positive (0).
3.2.7 Latch (holding) of the displayed value
• In a single-run measurement, etc., GO/NG judgment and analog output will be continued
while the measured data is latched (held) on the display for the specified period of time.
After the set period elapses, system operation returns to the ready state.
• Set up the display latch timer in the basic setup.
• While the display is being latched, inputs from the I/O interface or RS-232C/GP-IB are
still valid.