2 - 6
No. 99MBC095A
PROG
An error may be displayed at this stage,
however, the display at the right is not actually
an error. Check the shutter of the Measuring
Unit.
For information about other errors that may result refer to Section 8.3, Error
Messages and Remedies.
2.4 Initializing the LSM-6200 Display Unit
After making sure that this unit is operating normally, initialize the Display Unit so it can
recognize the Measuring Unit(s) to be used.
Initialization of the Display Unit is also required if the Measuring Unit needs to be changed.
In addition to replacing the ID unit that is associated with the Measuring Unit, initialize the
Display Unit (i.e. restore the factory setups) with the following procedure.
The initialization procedure is as follows:
Step 1: Turn off the power and connect the Measuring Unit with the ID unit that comes with
the Measuring Unit installed.
Step 2: Turn on the power while holding down the
C
key.
Hold down the
C
key for approximately 2 seconds, even after the power is on.
PROG
Step 3: When the self check has been completed, the
display shown at the right will appear. To initial-
ize, press the
ENT
key. When the initialization
process has been completed, the display restors
the initial conditions that existed just after the
power on.
To abort initialization press a key other than the
ENT
key or turn the power off.
In the former case the initialization process will be aborted and the initial display at
power-on will be restored.
NOTE Initialization will clear all the customer setup data and will restore the factory-setups.
Customize the setups again as necessary.