6 - 43
No. 99MBC095A
6. INTERFACE UNIT
3) Output signal
External power supply: Vcc
(5~24 V)
Internal GND External GND
SW1: Circuit 3
Internal circuit
R
Positive-true logic output by open collector
Voltage: 30V max., Drawing current: 10 mA max.
External resistance R: (Vcc /10) K R 10 k
Pin No. Signal name Remarks
Output of 7-digit measurement data (possible to select whether by
positive-true logic or by negative-true logic)
Output of segment error (positive-true logic)
1~28
29
DATA
ERR-0
Output of strobe (acknowledgment) (negative-true logic)
32 STB
Output of sign
HIGH level: Positive (+) data
LOW level: Negative (-) data
34 +POLE
4) Input signal
External power supply: Vcc
(5~24 V)
Internal power +5V
Internal GND External GND
F/R, HOLD
SW1: Circuit 3
SW1: Circuit 2
10K
Internal circuit
Pin No. Signal name Remarks
Does not hold the data when the circuit is open or the line level falls to LOW.
When the line level rises to HIGH, the previous data will be held (latched)
and the update does not take place. In order to read the data at a desired
point in time irrespective of the STB signal, start the read at least 500 µs after
the input of HOLD signal.
This is valid in simultaneous measurement.
Selects F (foreground program No.) when the circuit is open or the line level
falls to LOW.
Selects R (background program No.) when the line level rises to HIGH.
30
31
HOLD
F/R