Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
114 Chapter 1
GPIB Command Dictionary
CBUFfer subsystem
CBUFfer subsystem
CBUF controls the CDMA data buffer. This system allows you to send
CDMA data through the CDMA generator. You can send all zeroes,
external data via the DATA IN connector, random data, or data
supplied via GPIB.
:DATA <48 or 72 character hex string>
:DATA?
These commands set/query the actual data string sent in the CDMA
data buffer. The command CGEN:DATA:SOUR ‘Data Buff’ in the
“CGENerator subsystem” on page 146 selects the data buffer as the
data source.
Syntax
CBUF:DATA <hex string> !48 or 72 character hex string
CBUF:DATA?
Screen/field equivalent
CBUF:DATA is controlled by the Data Source field of the CDMA
GENERATOR screen, but :DATA has no direct screen equivalent.
:FRAMe:COUNt <integer> <real number>
:FRAMe:COUNt?
These commands set/query the number of frames to be transmitted. It
is applicable no matter what the source of data is (data buffer, zeroes,
external, or random). This command utilizes the “Multiple Real
Number Setting Syntax” on page 282, but does not use the :DUNits
and :UNITs commands.
Syntax
CBUF:FRAM:COUN <integer>,<real number>
CBUF:FRAM:COUN? <integer> !returns the count corresponding to
‘integer’
Screen/field equivalent
CBUF:FRAM:COUN controls the # of Frames field of the CDMA
GENERATOR screen. The # of Frame field is displayed when the Data
Source is set to Data Buff.