Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
222 Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CANalyzer:PNOFfset?
This command queries the PN offset measurement.
Syntax
MEAS:CAN:PNOF? !returns the measurement
Screen/field equivalent
MEAS:CAN:PNOF? reads the PN Offset measurement field on the
CDMA ANALYZER screen. This measurement is also displayed on the
CDMA GENERATOR screen.
:CANalyzer:POWer:ACP:LRATio <meas cmnd>
:CANalyzer:POWer:ACP:LRATio?
These commands set/query lower ratio measurement of the adjacent
channel power measurement. This ratio is the ratio of the total power in
the specified bandwidth at the specified offset below the center
frequency, to the power at the center frequency.
Syntax
MEAS:CAN:POW:ACP:LRAT <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CAN:POW:ACP:LRAT?
Screen/field equivalent
MEAS:CAN:POW:ACP:LRAT? reads the Lower ACP Ratio
measurement field on the CDMA ANALYZER screen when an ACP
measurement is selected. This measurement is also displayed on the
CDMA GENERATOR screen.