Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 185
GPIB Command Dictionary
CONFigure subsystem
:SPOR11|SPORT11|SB11|SP11:STOP <string>
:SPOR11|SPORT11|SB11|SP11:STOP?
These commands set/query the number of stop bits used for serial
communication over the SERIAL 11 port.
Syntax
CONF:SPOR11:STOP ‘1 bit’
‘2 bits’
CONF:SPOR11:STOP?
Screen/field equivalent
CONF:SPOR11:STOP controls the Stop Length field of the I/O
CONFIGURE screen.
:SPOR11|SPORT11|SB11|SP11:FCONtrol|FLOW <string>
:SPOR11|SPORT11|SB11|SP11:FCONtrol|FLOW?
These commands set/query the flow control to set the pace of serial
communications for the SERIAL 11 port.
Syntax
CONF:SPOR11:FCON ‘Xon/Xoff’
‘None’
CONF:SPOR11:FCON?
Screen/field equivalent
CONF:SPOR11:FCON controls the Flow Control field of the I/O
CONFIGURE screen.
:SPOR11|SPORT11|SB11|SP11:STATus:LINE?
This command queries the line status register.
Syntax
CONF:SPOR11:STAT:LINE?
Screen/field equivalent
No screen/field equivalent.