Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
146 Chapter 1
GPIB Command Dictionary
CGENerator subsystem
CGENerator subsystem
:CONTrol <string>
:CONTrol?
These commands set/query the CDMA generator’s mode. It defines the
type of data being sent.
Syntax
CGEN:CONT ‘Data’
‘Eb/No’
‘Noise’
CGEN:CONT?
Screen/field equivalent
CGEN:CONT controls the Gen Mode field of the CDMA GENERATOR
screen.
:DATA:SOURce <string>
:DATA:SOURce?
These commands set/query the type or source of the data being sent.
Syntax
CGEN:DATA:SOUR ‘Zeroes’ !Data is all zeroes
‘Ext’ !Data comes from the DATA IN port
‘Random’ !Generates random data
‘Data Buff’ !Gets data from the data buffer
CGEN:DATA:SOUR?
Screen/field equivalent
CGEN:DATA:SOUR controls the Data Source field of the CDMA
GENERATOR screen.