Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 119
GPIB Command Dictionary
CDANalyzer subsystem
:CPOWer:CALibrate
This command initiates the channel power calibration. When this
command is received, an internally generated calibration signal is
measured using the average power technique. Channel power is also
measured and a correction factor is generated. This correction factor is
applied to subsequent channel power measurements. Calibration
should be performed whenever a new set of measurements is made and
whenever the frequency of the measured signal is changed.
Syntax
CDAN:CPOW:CAL
Screen/field equivalent
CDAN:CPOW:CAL controls Chn Pwr Cal measurement field on the CD
Setup menu of the CODE DOM screen, when the CDMA standard is
set to “IS-95 Only”. The field displayed on the CD Setup menu when
the Measurement field on the Main menu is set to Power or Fast
Power, and the CD pwr unit field on the CD Setup menu is set to Abs.
:EVENt:QUALifier <string>
:EVENt:QUALifier?
These commands set/query the qualifying event for the trigger. The
qualifying event is the event that must occur before a trigger is
accepted. The trigger event is set by the :EVEN:TRIG command.
Syntax
CDAN:EVEN:QUAL ‘None’
‘27 ms’
‘20 ms’
‘80 ms’
‘2 s’
‘Ampl Lo’
‘Ampl Mid’
‘Ampl Hi
‘External’
Screen/field equivalent
CDAN:EVEN:QUAL controls the Qual Event field of the CODE DOM
screen. This field is displayed on the Trigger menu, when the CDMA
standard is set to “IS-95 Only”.