GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:MARKer:COMPlex:I[:LEVel]?
IS-2000 only. This query returns the Marker Level for the I channel
(Top bar), when the Complex Power Measurement is active. The return
is a floating point number.
Syntax
MEAS:CDAN:IS2000:MARK:COMP:I[:LEV]?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:COMP:I[LEV]? reads the I Pwr Lvl field
on the Main menu of the CODE DOMAIN screen when the CDMA
standard is set to “IS-2000” and the Measurement field is set to
Complex Pwr.
:CDANalyzer:IS2000:MARKer:COMPlex:Q[:LEVel]?
IS-2000 only. This query returns the Marker Level for the Q channel
(bottom bar) when the Complex Power Measurement is active. The
return is a floating point number.
Syntax
MEAS:CDAN:IS2000:MARK:COMP:Q[:LEV]?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:COMP:Q[:LEV]? reads the Q Pwr Lvl
field on the Main menu of the CODE DOMAIN screen when the CDMA
standard is set to “IS-2000” and measurement is Complex Pwr.
:CDANalyzer:IS2000:MARKer:LEVel:ABSolute?
IS-2000 only. This query returns the Marker Level, in dBm, when the
Channel power mode is active. The return is a floating point number
formatted as shown on the screen.
Syntax
MEAS:CDAN:IS2000:MARK:LEV:ABS?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:LEV:ABS? reads the Marker Lvl field of
the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000” and CD pwr unit is set to Abs.