292 Chapter 1
GPIB Command Dictionary
OSCilloscope subsystem
:TRIGger:LEVel <real number>
:TRIGger:LEVel?
These commands set/query the level of the oscilloscope’s trigger.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STAT, :DUN, :UNIT, :INCR:MODE, :INCR:DUN,
commands.
Syntax
OSC:TRIG:LEV <real> !values
OSC:TRIG:LEV? !returns the level
Example
OSC:TRIG:LEV 0.4 !sets the level to 0.4 V
Screen/field equivalent
OSC:TRIG:LEV controls the upper subfield of the Level(div) field on
the Trigger menu of the SCOPE screen.
:TRIGger:MODE <string>
:TRIGger:MODE?
These commands set/query the retriggering mode of the trigger for the
oscilloscope. If ‘Cont’ is chosen, then the oscilloscope is triggering
continuously and is ready for another measurement after one has
finished. If ‘Single’ is chosen then the oscilloscope requires a trigger
command before making another measurement.
OSC:RESet is the trigger command for the oscilloscope.
Syntax
OSC:TRIG:MODE ‘Cont’
‘Single’
OSC:TRIG:MODE?
Example
OSC:TRIG:MODE ‘Single’ !sets mode to single trigger
Screen/field equivalent
OSC:TRIG:MODE controls the Cont/Single field on the Trigger menu
of the SCOPE screen.