Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 321
GPIB Command Dictionary
SANalyzer subsystem
:MASK:BEEP
:MASK:BEEP?
This command turns the failure indicator beeper off or on.
Syntax
SAN:MASK:BEEP ‘On’
‘Off’
SAN:MASK:BEEP?
Screen/field equivalent
SAN:MASK:BEEP controls the Mask Beep field on the Mask menu of
the SPEC ANL screen.
:MASK:DISPlay
:MASK:DISPlay?
This command sets/queries which masks will be displayed: upper,
lower, both, or none (Off).
Syntax
SAN:MASK:DISP ‘UpperOnly’
‘LowerOnly’
‘Both’
‘Off’
SAN:MASK:DISP?
Screen/field equivalent
SAN:MASK:DISP controls the Display field on the Mask menu of the
SPEC ANL screen.
:MASK:FIXed:LOWer:POINts:NUMBer
:MASK:FIXed:LOWer:POINts:NUMBer?
This command sets/queries the number of points in the lower spectrum
analyzer mask. The maximum number is 15.
Syntax
SAN:MASK:FIX:LOW:POIN:NUMB <integer 1-15>
SAN:MASK:FIX:LOW:POIN:NUMB?
Screen/field equivalent
SAN:MASK:FIX:LOW:POIN:NUMB controls the #Pts field on the Mask
menu of the SPEC ANL screen when the Edit Mask field is set to
Lower, and the Mask Type field is set to Fix.