Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 99
GPIB Command Dictionary
CANalyzer subsystem
CANalyzer subsystem
The CANalyzer subsystem controls functions unique to the CDMA
ANALYZER screen.
:ACP:POWer:CALibrate
This command initiates calibration of the adjacent channel power
measurement for the current RF analyzer frequency at the selected
frequency offset and filter bandwidth.
Syntax
CAN:ACP:POW:CAL
Example
CAN:ACP:POW:CAL !calibrates ACP measurement
Screen/field equivalent
CAN:ACP:POW CAL controls the ACP Cal field on the CDMA
ANALYZER screen.
:ACP:POWer:FILTer <real number>
:ACP:POWer:FILTer?
This command selects the filter bandwidth over which the adjacent
channel power measurement is made in the adjacent channels. Valid
values are 10 kHz to 1.23 MHz with 500 Hz resolution.
Syntax
CAN:ACP:POW:FILT <real number>
CAN:ACP:POW:FILT?
Example
CAN:ACP:POW:FILT 10 kHz !sets the filter BW to 10 kHz
CAN:ACP:POW:FILT? !returns the filter BW setting
Screen/field equivalent
CAN:ACP:POW:FILT controls the ACP Fltr BW field on the CDMA
ANALYZER screen.