Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 209
GPIB Command Dictionary
I/Q Plot Subsystem
:TRIGger:DARM
This command disarms the IQ Plot measurement when it has been put
into single triggering mode using the TRIG:MODE:RETR SING. See
“TRIGger subsystem” on page 346 for a more complete discussion of
GPIB triggering.
Syntax
IQPL:TRIG:DARM !disarms the IQ Plot measurements.
Screen/field equivalent
IQPL:TRIG:DARM controls the Disarm selection of the Analyzer field
of the Main image of the IQ PLOT screen.
:TRIGger:DELay <real number>
:TRIGger:DELay?
These commands set/query the current field contents of the
Trig Delay field on the IQ Plot screen.
Syntax
IQPL:TRIG:DEL <real number from 20µs to 10s>
IQPL:TRIG:DEL?
Screen/field equivalent
IQPL:TRIG:DEL sets the Trig Delay field of the Trigger image when
the Trig Event field is set to Delay on the IQ PLOT screen. The query
returns the current value.
:TRIGger[:EVENt] <string>
:TRIGger[:EVENt]?
These commands set/query the trigger interval in the IQ plot screen.
Syntax
IQPL:TRIG[:EVEN] ‘27 ms’
‘20 ms’
‘80 ms’
‘2 s’
‘Delay’
‘Immed’
IQPL:TRIG[:EVEN]?
Screen/field equivalent
IQPL:TRIG[:EVEN] sets the Trig Event field of the Trigger image
on the IQ PLOT screen.