Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
236 Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:MARKer:CHANnel:RATE:RC3?
IS-2000 only. This query returns the RC-3 data rate value for the
channel selected by the Marker’s current position. The return value is
an integer.
Syntax
MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC3?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC3? reads the RATE RC-3
field of the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000” and Sprd Fact is selected.
:CDANalyzer:IS2000:MARKer:CHANnel:RATE:RC4?
IS-2000 only. This query returns the RC-4 data rate value for the
channel selected by the Marker’s current position. The return value is
an integer.
Syntax
MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC4?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC4? reads the RATE RC-4
field of the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000” and Sprd Fact is selected.