224 Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CANalyzer:POWer:AVG <meas cmnd>
:CANalyzer:POWer:AVG?
These commands set/query the average power measurement.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
Syntax
MEAS:CAN:POW:AVG <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CAN:POW:AVG?
Example
MEAS:CAN:POW:AVG? !returns the measurement
Screen/field equivalent
MEAS:CAN:POW:AVG? reads the Avg Pwr measurement field on the
CDMA ANALYZER screen. This measurement is also displayed on the
CDMA GENERATOR screen.
:CANalyzer:POWer:CHANnel <meas cmnd>
:CANalyzer:POWer:CHANnel?
These commands set/query the channel power measurement.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
Syntax
MEAS:CAN:POW:CHAN <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CAN:POW:CHAN?
Example
MEAS:CAN:POW:CHAN? !displays the measurement
Screen/field equivalent
MEAS:CAN:POW:CHAN? reads the Chan Pwr measurement field on
the CDMA ANALYZER screen. This measurement is also displayed on
the CDMA GENERATOR screen.